Flow-Through Cells with Optical Temperature Sensor

  • Flow-through cells with integrated Optical Temperature Sensor
  • Luer-lock with tubing adapters
  • For measurements in gas and water
  • >3 years shelf time 

The new optical temperature sensors from PyroScience are based on an innovative technology utilizing temperature-sensing properties of luminescent indicators. The flow-through cells have an integrated optical temperature sensor for contactless measurements in gaseous or liquid samples flowing through the cell. They are available as small version TPFTC and as large version TPFTC2. The cell is equipped with Luer-lock connectors on both tubing ends. The included set of Luer-lock adapters allows direct connection of tubings with ca. 1-2 mm inner diameter (TPFTC) or ca. 3-4 mm inner diameter (TPFTC2).

Gallery

Compatible oxygen meters

FireStingO2 (FSO2-x)
FireStingO2-Mini (FSO2-MINI-ST)
Piccolo2 (PICO2, PICO2-OEM)

Available Items

Item No. Description Pricing
in EU*
TPFTC2 Flow-Through Cell with optical Temperature Sensor (2 pieces per package), incl. Luer-lock adapters for ca. 3-4 mm ID tubings, recommended flow rate: 20-500 mL/min 200 €
TPFTC Flow-Through Cell with optical Temperature Sensor (2 pieces per package), incl. Luer-lock adapters for ca. 1-2 mm ID tubings, recommended flow rate: 10-100 mL/min 200 €

Accessories for Connection to FireSting Devices

Item No. Description Pricing
in EU*
SPFIB-BARE Bare Optical Fiber (1m length), ST-plug + one polished end 50 €
SPFIB-BARE-CL2 Bare Optical Fiber (2m length), ST-plug + one polished end 60 €

Accessories for Connection to Piccolo2 Devices

Item No. Description Pricing
in EU*
PICFIB2 Optical Fiber (1m length) for Piccolo2, bare polished end 70 €

* Only EU-Prices shown. Contact us for pricing outside EU. Not for resale.

Specifications

All versions
Measuring range 0 to 50°C
273-323 K
32-122 °F
Resolution 0.02°C
Accuracy +/- 0.5°C (1-point calibrated)
Drift < 0.5°C / day
Minimum lifetime 10,000,000 data points
Application areas Laboratory, industry, research. Not for medical or any other safety-critical applications. Not for application in humans or in food intended for human consumption.
Compatibility Gas-phase, water/aqueous solutions
Cross sensitivity Organic solvents
Cleaning 3 % H2O2, ethanol, soap solution
Storage time >3 years in darkness at room temperature
Calibration 1-point calibration
Tubing connector Luer-lock with adapters for ca 1-2 mm ID tubings (TPFTC) or 3-4 mm ID tubings (TPFTC2)